Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 97
Operation
7.3.1.1 Manual ventilation
► Set the APL valve to the desired maximum airway pressure.
The patient can be ventilated with the breathing bag. The pressure is limited to the
set value.
In the Manual / Spontaneous mode, lifting the valve relieves pressure from the
breathing system.
7.3.1.2 Spontaneous breathing
► Turn the APL valve counterclockwise as far as it will go.
The dots are aligned vertically over one another. The valve lifts.
The pressure limitation is canceled and the valve is open for free spontaneous
breathing.
7.3.2 Using the O2 flush
The O2 flush is used for flushing and quickly filling the breathing system and
breathing bag with oxygen. The vaporizers are bypassed for this.
► Press the O
2+ key.
O
2 continues to flow for as long as the key is pressed.
3525635264
20 3010