Maintenance
Fast Response Mode
In the factory-preset configuration ("normal response mode"), the gas transmitter provides a
minimum measured value response time of 4 seconds. The gas transmitter can be switched to
"fast response mode" using the Dräger CC-Vision GDS Software or (optional) with a software
based on HART
®
DD/DTM or a HART
®
handheld terminal. This can be useful when the
measured gas is fed via pump or when the gas transmitter is operated in ducts where fast
monitoring of gas concentration changes is required.
Due to the system design, very fast gas changes could lead to increased measured values.
Remedy (if necessary): Reduce the gas exchange rate by reducing the flow rate at the process
cuvette or at the process adapter, and/or in the duct.
Disposing of the Device
CAUTION
When "fast response mode" is set, the repeatability error increases by the factor 2. After
changing the response mode, it is absolutely necessary to perform a complete calibration.
From August 2005, regulations apply to the whole of the EC for the disposal of
electrical and electronic devices; these regulations are set in the EU Directive 2002/
96/EC and in national laws relating to this device.
Special collection and recycling possibilities have been established for private
households. However, since the device has not been registered for use in private
households, it cannot be disposed of as special household waste. For disposal, it can
be sent back to your national Dräger Safety Sales organisation; if you have any
questions requiring disposal, then please contact this location.

Table of Contents

Other manuals for Dräger PIR 7000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger PIR 7000 and is the answer not in the manual?

Dräger PIR 7000 Specifications

General IconGeneral
Detection PrincipleInfrared (IR)
Pressure Range700 to 1300 hPa
Ingress ProtectionIP66/67
Gases DetectedHydrocarbons
Measurement Range0 - 100 %LEL (Lower Explosive Limit) or 0 - 100 vol.%
Operating Temperature-40 to +65 °C
Output Signal4-20 mA
Housing MaterialStainless steel
CertificationsATEX, IECEx, UL, CSA
Response Timet90 ≤ 4 s