Accessories
Insect Guard PIR 7000
*)
(Part no. 68 11 609)
Hydrophobic Filter PIR 7000
**)
(Part no. 68 11 890)
Intended use:
The hydrophobic filter is designed to protect the measuring cuvette in applications with
extreme dust or dirt accumulation.
Dust and dirt particles can settle on the optical surfaces of the gas transmitter thereby
resulting in an early malfunction.
The hydrophobic filter must only be used in combination with status indicator PIR 7000 /
7200 (part no. 6811625/6811920), the flowcell PIR 7000 / 7200 (part no. 6811490/
6811910) or the bump test adapter PIR 7000 / 7200 (part no. 6811630/6811930).
*)
The insect guard PIR 7000 is also suitable for the use with Dräger PIR 7200.
Intended use:
The insect guard is designed to prevent
insects from blocking the gas inlet and
outlet apertures as well as the inner gas
passages in the splash guard.
Mounting:
Slot both parts of the insect guard into
the openings of the splash guard.
NOTICE
Using the insect guard lengthens the
response time of the gas transmitter, see
"Performance Characteristics (typical
values)" on page 73.
**)
The hydrophobic filter PIR 7000 is also suitable for the use with Dräger PIR 7200.
01523886_01.eps

Table of Contents

Other manuals for Dräger PIR 7000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger PIR 7000 and is the answer not in the manual?

Dräger PIR 7000 Specifications

General IconGeneral
Detection PrincipleInfrared (IR)
Pressure Range700 to 1300 hPa
Ingress ProtectionIP66/67
Gases DetectedHydrocarbons
Measurement Range0 - 100 %LEL (Lower Explosive Limit) or 0 - 100 vol.%
Operating Temperature-40 to +65 °C
Output Signal4-20 mA
Housing MaterialStainless steel
CertificationsATEX, IECEx, UL, CSA
Response Timet90 ≤ 4 s