73
Technical Data
Performance Characteristics (typical values)
Dräger PIR 7200
Carbon dioxide
±0.01 % vol.
±0.05 % vol.
±0.02 % vol.
±0.1 % vol.
±0.5 % vol.
±1.5 % vol.
±4.5 % vol.
±40.0 % vol.
±0.03 % vol.
±0.1 % vol.
±0.003 % vol./
o
C
±0.012 % vol./
o
C
±0.035 % vol./
o
C
±0.25 % vol./
o
C
±0.005 % vol.
±0.15 % vol.
0.16 %/mbar
* based on LEL values acc. to IEC
** for calibration at 50% FSD
Dräger PIR 7000 Type 340
Methane
±0.7 %LEL
±1.0 %LEL
±1.4 %LEL
±2.0 %LEL
±2.5 %LEL
±2.0 %LEL
±3.0 %LEL
±0.12 %LEL/
o
C
±1.5 %LEL
±3.1 %LEL
0.16 %/mbar
Propane
±0.25 %LEL
±0.25 %LEL
±0.5 %LEL
±0.5 %LEL
±1.0 %LEL
±0.6 %LEL
±1.0 %LEL
±0.07 %LEL/
o
C
±0.8 %LEL
±1.1 %LEL
0.13 %/mbar
Dräger PIR 7000 Type 334
Ethylene
±1.0 %LEL
±1.0 %LEL
±2.0 %LEL
±2.0 %LEL
±2.4 %LEL
±2.0 %LEL
±3.0 %LEL
±0.13 %LEL/
o
C
±1.7 %LEL
±1.2 %LEL
0.16 %/mbar
Propane
±0.25 %LEL
±0.25 %LEL
±0.5 %LEL
±0.5 %LEL
±1.2 %LEL
±1.0 %LEL
±2.0 %LEL
±0.1 %LEL/
o
C
±0.5 %LEL
±0.9 %LEL
0.13 %/mbar
Methane
±0.5 %LEL
±0.5 %LEL
±1.0 %LEL
±1.0 %LEL
±1.5 %LEL
±1.0 %LEL
±1.0 %LEL
±0.17 %LEL/
o
C
±0.5 %LEL
±2.4 %LEL
0.16 %/mbar
Repeatability
in "normal response mode":
zero-point
50 %LEL*
5 % vol.
in "fast response mode":
zero-point
50 %LEL*
5 % vol.
Linearity error (max.), for
measuring range **:
0 - 100 %LEL
0 - 5 % vol.
0 - 10 % vol.
0 - 30 % vol.
0 - 100 % vol.
Long-term drift (12 months):
zero-point
Temperature influence,
–40 to 77
o
C:
zero-point
50 %LEL*
1 % vol.
5 % vol.
10 % vol.
30 % vol.
Humidity influence, 0 to 100 %
rel. humidity at 40
o
C:
zero-point
50 %LEL*
5 % vol.
Pressure influence:

Table of Contents

Other manuals for Dräger PIR 7000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger PIR 7000 and is the answer not in the manual?

Dräger PIR 7000 Specifications

General IconGeneral
Detection PrincipleInfrared (IR)
Pressure Range700 to 1300 hPa
Ingress ProtectionIP66/67
Gases DetectedHydrocarbons
Measurement Range0 - 100 %LEL (Lower Explosive Limit) or 0 - 100 vol.%
Operating Temperature-40 to +65 °C
Output Signal4-20 mA
Housing MaterialStainless steel
CertificationsATEX, IECEx, UL, CSA
Response Timet90 ≤ 4 s