EasyManuals Logo

Dräger PIR 7000 Technical Handbook

Dräger PIR 7000
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #55 background imageLoading...
Page #55 background image
55
Maintenance
Example: If the gas transmitter Dräger PIR 7000 has been calibrated under normal air
pressure (1013 mbar), a pressure change to 900 mbar would result in the following change of
the measuring value:
(Remedy: Applying 50 %LEL calibration gas at 900 mbar pressure)
Preparing Calibration
Provide calibration gas.
For gas transmitters with mounted splash guard PIR 7000 / 7200:
Push calibration adapter PIR 7000 onto the splash guard until it locks into place.
For gas transmitters with mounted flowcell PIR 7000 / 7200, bump test adapter PIR 7000 /
7200, process adapter PIR 7000 or process cuvette PIR 7000:
Connect calibration gas feed directly to the flowcell, the bump test adapter, the process
adapter or the process cuvette.
Calibration Using the Magnetic Wand
The calibration can be performed directly at the gas transmitter using the magnetic wand (part
no. 45 43 428).
50 %LEL X
900 mbar
= 44 %LEL
1013 mbar
NOTICE
When delivered, the gas transmitter is calibrated with a reference pressure of 1013 mbar.
Deviations from this reference pressure can lead to deviating measured values at any
concentration.
CAUTION
Always first calibrate the zero-point before the sensitivity (span).
Calibration using the magnetic wand:
The gas transmitter has the marks on two
opposite sides; » 0 « for "Zero-point" and
» S « for "Sensitivity (span)".
For calibration, place the magnetic wand
onto the respective mark.
03023886_01.eps
S
0

Table of Contents

Other manuals for Dräger PIR 7000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger PIR 7000 and is the answer not in the manual?

Dräger PIR 7000 Specifications

General IconGeneral
Detection PrincipleInfrared (IR)
Pressure Range700 to 1300 hPa
Ingress ProtectionIP66/67
Gases DetectedHydrocarbons
Measurement Range0 - 100 %LEL (Lower Explosive Limit) or 0 - 100 vol.%
Operating Temperature-40 to +65 °C
Output Signal4-20 mA
Housing MaterialStainless steel
CertificationsATEX, IECEx, UL, CSA
Response Timet90 ≤ 4 s

Related product manuals