XS series operating manual
XS user manual v0.55 - 127 -
Continuity : VOLT 2.830E+00 OHM 3.230E-1 (main unit =VOLT)
B) Following characters in sequence mode
For instance :
L1 C0: 0.15m 0.00V,L2 R0: 1.50KV 0.02mA,L3 M0: 41.7 G,L4 F0:q 0.01mA 223V
A2,L5 ..: ,L6 ..: ,L7 ..: ,L8 ..:
The syntax rules are the following :
the end of the message is : CR ( Carriage Return)
each test step starts with a « L » and ends with « , » ( except for the last one) .
the « L » letter is followed by a number between 1 and 8 , giving the test step number and
then by a « space ».
then comes a letter according to the type of the test step :
the measurement type is followed by the memory number which contains the
measurement parameters , then by a separator character « : »
if the test step is void , the measurement will be replaced by a « space »
otherwise , the next character gives the test step result :
« space » indicates that the test is good
« q » indicates that the test is bad
for the insulation and dielectric strength tests the previous character is followed by 1
‘space’ and for the ground continuity test by 2 ‘space’.
then the numerical values are given . When the value cannot be measured ( no continuity
or overload ) the numerical indication is replaced by « ---- »
To check if a tested specimen is good or bad at the end of a sequence , it is necessary to
check the STB register ( bit 3) .To get the result of each test step , check if the result
character is « space » or « q ».
RS232 / IEEE-488 commands summary