XS series operating manual 
 
XS user manual v0.55  - 48 - 
 
Example of a sequence setup: 
An operator wants to perform an insulation control on a specimen, then a dielectric strength test 
and finally a new insulation control to verify that the dielectric strength test hasn't degraded the 
specimen. 
  SETUP menu 
  <page2>, line SEQUENCE 
  Select the sequence mode : M+H+M 
  Escape with [ESC] key 
  The unit switches automatically in the insulation test 
  Check that parameters are correct (Necessary to select times) 
  Press on the [MEASUREMENT-DISCHARGE] push button 
  After the test time the unit switches automatically to the next test if the test result was 
PASS. 
  When  an  insulation  test  has  to  be  perform  after  a  dielectric  strength  test,  the  unit 
controls the  discharge voltage  of the  specimen  and follows its sequence  only if  the 
voltage is lower than 100 volts with the display of : 
 
    DISCHARGE CONTROL 
 
In case of a fault during the insulation test, the number 1 or 2 of the test is displayed. 
At  the  end  of  a  sequence,  after  release  of  the  result  by  pressing  the  [MEASUREMENT-
DISCHARGE] push button the unit switches in the first function of the sequence. 
To escape of the SETUP menu, press on the function key [ESC]. 
 
 
[expert] 
EXPERT mode Activation  
 
 
Fig 6.11 
 
The EXPERT mode gives additional controls for function such as the dielectric strength test  
( breakdown detection mode and  breakdown threshold values during the rise time ) 
This mode is only for skill operators. 
Any information related to this operating mode is indicated with this label: [expert]