Trouble shooting the RS232 interface 
 
When  the  RS232C  interface  is  not  operating  as  described  in  this  manual,  please  check  the 
following points: 
 
No reaction from the XS unit when sending commands : 
 
o  The cable between the XS unit and the PC computer must be correctly connected at both 
ends. The cable is a special wiring cable : check that the cable is a Sefelec CO179 model or 
check that the wiring has been done according to the instructions . 
 
o  The XS series units have several possible interface types : check that in the SETUP menu 
<page2>, the RS232 mode has been selected on the INTERFACE line. 
 
o  The end of message must be the LF (hexa 0A,decimal 10) character. If this character is not 
sent,  the  XS  unit  won’t  handle  the  message  :  check  that  this  character  is  added  to  the 
command. 
 
o  The first command must be the REM command, which displays the REMLOC message on 
the LCD display, the unit being ready to receive all the others commands.  
 
o  WARNING : the first REM command should not wait for the Xon character before being sent. 
 
 
 The unit goes in REMOTE mode and then hang up : 
 
Before sending a new command, it is mandatory to wait that the XS unit indicates its non busy 
state by sending the Xon character (hexa 11,decimal 17). If a code arrives during the handle of 
the previous one, it can produce an erratic operating mode of the unit or stop it.