Length of power failure Control mode Behaviour of pump
Power is restored aer the rot
a-
onal speed of the pump decrea-
ses below 50% but before regen-
erave power stops
Parallel or serial control mode,
mer disabled
Regenerave power keeps all
output signals acve during the
power failure. As the mer is dis-
abled, the controller will go into
fail condion when the speed
decreases below 50% and will
display ashing error code 0.
When the power is restored, the
pump will not speed up unl the
error is cleared. To clear the er-
ror, send a stop command (paral-
lel or serial, depend on the con-
trol mode), send a start com-
mand to speed up the pump to
the designated speed.
Parallel or serial control mode,
mer enabled
Regenerave power keeps all
output signals acve during the
power failure. If power is re-
stored before the mer period
expires, the pump will speed up
to designated speed. If the mer
period expires, the controller will
go into fail condion and will dis-
play ashing error code 3. As
above, when the power is re-
stored, the pump will not speed
up unl the error is cleared. To
clear the error, send a stop com-
mand (parallel or serial, depend
on the control mode), send a
start command to speed up the
pump to the designated speed.
Power is restored aer the rot
a-
onal speed of the pump decrea-
ses below 50% and the regenera-
ve power stops
Parallel or serial control mode
Regenerave pow
er stops and
fails to keep output signals.
When power is restored, the
pump with parallel control will
automacally restart if the start
control on the logic interface is
set to start. The pump with serial
control requires a new start com-
mand to speed up the pump to
the designated speed. The fail
signals caused during the regen-
erave power period are lost
when the power is restored.
04/2022 - ©Edwards Limited
Page 79B80000880_G
B80000880_G - Operaon