Table 25
Flashing service codes
Service ash
code
Comments Acons
LED on 1sec
LED o 1sec
Oil cartridge service rec-
ommended
We recommend you to replace the oil car-
tridge. R
efer to Service on page 95.
LED on 3sec
LED o 1sec
Bearing and oil car-
tridge ser
vice recom-
mended
We recommend you to replace the bear-
ing and oil cartridge. Refer to Service on
page 95.
LED on 3sec
LED o 3sec
Pump service necessary
The turbo impeller or controller has
reached its expected life. We recommend
you to return the pump to our service
centre for service. Refer to Service on
page 95.
8.2 Decoding service status words
The service status can be accessed through the serial link. This method of accessing
service status will give the most complete informaon of current and future service
requirements and preventave maintenance acvies can be scheduled.
A summary of the current pending service status is given in response to the service
status command that follows:
Command ? V 8 8 1 cr
The reply is as follows:
Reply = V 8 8 1 sp h h h h h h h h cr
The service status word is made up of 8 hexidecimal digits. To decode this word, convert
each digit int
o a 4‑digit binary number as described in Decoding system status words on
page 89.
Each binary digit (bit) represents a ag that is acve (state 1) or not acve (state 0). To
help decode the service status word, each bit is numbered (starng with 0 for the least
signicant to 31 for the most signicant) as shown in Controller run me on page 87. The
meaning of each bit in the service status word is given in Table: Service ags.
Table 26
Service ags
Bit number
Status ag Acve ag means
0 Oil cartridge service due
Set when hours unl oil cartridg
e service due
= 0
1 Bearing service due Set when hours unl bearing service due = 0
2 Pump ser
vice due
Set when hours unl pump service due = 0
or cy
cles unl pump service due = 0
3 Controller service due
Set when hours unl controller service due =
0
4 Reserved -
5 Reserved -
04/2022 - ©Edwards Limited
Page 86B80000880_G