EasyManuals Logo

Edwards nEXT400 User Manual

Edwards nEXT400
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #28 background imageLoading...
Page #28 background image
Figure 8 Max allowed rate of pressure rise during venng: pr
essure against me (pump inially at
full speed)
3.4 Purge gas specic
aon
Table 8
Purge gas specicaon
Purge gas specicaon Reference data
Purge gas Dry air, nitrogen, argon or other inert
gases
Maximum dew point at an atmospheric
pressure
-22 °C
Maximum size of parculates 1 µm
Maximum concentraon of oil 0.1 parts per million
Permied purge gas ow (when necessary)
20 to 50 sccm (0.33 to 0.84 mbar l s
-1
or
33 to 84 P
a l s
-1
)
Recommended purge gas ow
25 sccm (0.42 mbar l s
-1
, 42 Pa l s
-1
)
Ma
ximum permied supply pressure of the
purge gas
1 bar (gauge), 14.5 psig, 2 x 10
5
Pa
3.5 Cooling w
ater
Re
fer to Table: Cooling water specicaon for the cooling water specicaon. The
cooling water supply must be a highquality drinking water specicaon. Check with the
water supply authority if there is doubt about the quality of the supply.
Table 9
Cooling water specicaon
Cooling water specicaon Reference data
Quality
Mechanically clean and opcally clear
with no deposits or turbidity
04/2022 - ©Edwards Limited
Page 28B80000880_G
B80000880_G - Technical data

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards nEXT400 and is the answer not in the manual?

Edwards nEXT400 Specifications

General IconGeneral
BrandEdwards
ModelnEXT400
CategoryWater Pump
LanguageEnglish

Related product manuals