SCU-801 Control Unit for Turbomolecular Pump
*1 The driver
output
stops and the rotor continue rotating by
inertia.
It may take several hours to stop
. To stop the STP pump quickly, close the vacuum valve at the outlet port flange
and introduce gas from the purge port into the STP pump.
Table
35 - Error
messages after self test
Check the vacuum piping
system for leakage. Check the
exhaust sequence.
1. Perform rotational
operation
after resetting.
2. Contact
Service office.
1. Set the ambient
temperatur
e to 40 ºC (104
ºF) or less.
2. Remove obstacle from the
ventilation port.
Atmospheric air flows the STP
pump
.
1. Malfunction by external
noise
.
2. Failure of the SCU
-801.
1. Disconnection of
connection cable or at
connector part
2. STP pump failure
.
Short
-circuit or ground fault in
the connection cable or at
connector part
.
1. High ambient temperature.
2. Insufficient cooling of the
SCU-801.
Decelerate and stop,
then Touch down
.
The magnetic
bearing output stops
and the rotor is
descended on the
protection bearing.
Touch down
. The
magnetic bearing
output stops and the
rotor is descended
on the protection
bearing.
Touch down
. The
magnetic bearing
output
stops and the
rotor is descended
on the protection
bearing.
Rotational speed rises
when deteriorating or
stopping.
When the rated operation
does not stop approx. 30
min. after start operation is
executed.
Accelerate up to 500
rpm
or less
.
Disconnection of magnetic
bearing control current.
Excessive magnetic
bearing electric current
.
Overheat inside the
SCU-801.