SCU-801 Control Unit for Turbomolecular Pump
*1 The driver output stops and the rotor continues rotating by inertia.
It may take several hours to stop. To stop the STP pump quickly, close the vacuum valve at the
outlet port flange
and introduce gas from the purge port into the STP pump.
Table
36 - Error
messages after self test (continued)
1. Set the ambient
temperature to 40 ºC (104
ºF) or less.
2. Remove obstacle from the
ventilation port.
1. When changing the cable
length, perform tuning.
2. Contact Service office.
1. Remove external vibration
so as not to transmit it to the
STP pump.
2. Check the vacuum piping.
3. Install the STP
pump in a
way that no foreign matter
falls into the S
TP pump.
1. High ambient temperature.
2. Insufficient cooling of the
SCU-801.
Abnormal magnetic bearing:
1. Change of the cable
length.
2. Disconnection of the STP
connection cable or control
circuit error.
Excessive vibration applied
externally to the STP
pump:
1. External vibration/impact.
2. Atmospheric air flows into
the STP
pump.
3. Foreign matter falls into the
STP pump.
Contact noise and
vibration of the rotor
are generated from
the STP
pump.
Decelerate and stop.
Overheat inside the
SCU-801.
Failure of serial
communication.
Abnormal driver operation
.
Disturbance X_B
Disturbance X_H
Disturbance Y_B
Disturbance Y_H
Disturbance Z