SCU-801 Control Unit for Turbomolecular Pump
Table
42 - Error
messages after self test (continued)
1. Connect the
electromagnetic cooling
water valve correctly.
Check the volume of water
and water temperature.
2. Contact
Service office.
1. Connect the
electromagnetic cooling
water valve correctly.
Check the volume of water
and water temperature.
2. Contact Service office.
1. Contact Service office.
2. When the TMS function is
not in use, set the TMS
function to "DISABLE"
.
3. Connect the cable
correctly. If the cable
connector is not locked,
lock it securely.
4. Contact Service office.
1. Perform tuning after
power
ON/OFF
operation.
2. When the error is displayed
after tuning, contact
Service office.
1. Cooling
failure of
TMS unit
.
2. Failure of
the TMS unit.
1. Cooling failure of TMS unit
.
2. Failure of the TMS unit.
1. Disconnection of the TMS
sensor.
2. Incorrect setting of the
TMS function.
3. Disconnection of the TMS
connection cable.
4. Failure of the TMS
connection cable.
1. Failure of the STP
pump.
2. Failure of the SCU
-801.
Touch down. The
magnetic bearing
output stops and the
rotor is descended
on the
touch down
bearing.
TMS temperature control
error (when the TMS unit
detection temperature
becomes higher than the
setting value by 10°C (18
ºF) or more.
TMS temperature control
error (when the TMS unit
detection temperature
becomes lower than the
setting value by 10°C
(18
ºF) or more.
Error is not
detected for a certain time
after the STP pump starts.
TMS sensor is not
connected when the state
"TMS Functi
on" is
"ENABLE"
, "External" or
"Advanced".
Tuning Error 1
:
:
Tuning Error
19