SEP 2005 MODEL 1000
INSTALLATION AND SETUP
3-24
(a)
EQUIPMENT DAMAGE OR PERSONAL INJURY
Do not apply AC electrical power to the Analyzer or the GC Controller until all electrical
power, interconnection, and external signal connections have been verified, and proper
grounds have been made. Refer to Section 3.1.3 for general precautions concerning
signal wiring.
Failure to properly connect the GC unit may result in serious equipment damage or personal
injury.
(6) Access the GC Controller's Terminal Board for Field Wiring (TB), and connect the other
leads of the Interconnect Cable to the GC Controller's TB (see instructions in Section 3.4.2).
Ensure that the connections correspond to those listed in Table 3-2 and Figure 3-3.
(a) At this time, you may also want to complete the remainder of all electrical
connections at the GC Controller. If so, see all of Section 3.4, this manual.