Figure A-5: Loop Resistance: Passive Non-IS Analog Output
%
$









$
A. Loop resistance [Ω]
B. External power supply voltage [V]
Max Loop Resistance
(14)
@ 23 mA = 43.4 * (External Power Supply Voltage – 8) [Ω]
Figure A-6: Loop Resistance: Passive IS Analog Output
%
$







$
A. Loop resistance [Ω]
B. External power supply voltage [V]
Max Loop Resistance
(14)
@ 23 mA = 43.4 * (External Power Supply Voltage – 9.4) [Ω]