Duplex 2205 (EN 1.4462) according to EN 1092-1 material group 16E0:
Standard: Max. 580 psig (40 Bar), -22 °F (-30 °C) up to max 302 °F (150 °C)
(14)
HP: PN 320, -22 °F (-30 °C) up to max 482 °F (250 °C)
(14)
HTHP: PN 320, -22 °F (-30 °C) up to max 482 °F (250 °C)
(14)
JIS flange rating
316 according to JIS B2220 material group 2.2:
Standard: Max. 302 °F/580 psig (150 °C/40 Bar)
HP: Max. temp. 250 °C. Final rating depends on flange.
C: Max. temp. 200 °C. Final rating depends on flange.
HTHP: Max. temp. 400 °C. Final rating depends on flange.
Fisher and Masoneilan flange rating
316 according to ASME B16.5 Table 2-2.2:
Standard: Max. 302 °F/580 psig (150 °C/40 Bar)
HP: Class 600 up to max 250 °C
C: Class 600 up to max 200 °C
HTHP: Class 600 up to max 400 °C
A.2.11
Tri-Clamp
®
rating
Tri-Clamp is available for the Standard Temperature and Pressure seal.
Table A-11: Tri-Clamp Rating
Size Maximum pressure (bar)
(1)
1½-in. (37.5 mm) 16
2-in. (50 mm) 16
3-in. (75 mm) 10
4-in. (100 mm) 10
(1) The final rating depends on the clamp and gasket.
A.2.12 Plate design
Certain models of flanged alloy and PTFE covered probes have a tank connection design
with a protective flange plate of the same material as the probe and with a backing flange
in 316L/EN 1.4404. The protective flange plate prevents the backing flange from being
exposed to the tank atmosphere.