Procedure
1. Select Setup → Tank.
2. Select the Environment tab.
3. Under Advanced select/deselect the Thin Layer Detection checkbox to enable/
disable the thin layer detection.
4. Select Store.
D.9.2 Enable/disable thin layer detection using AMS Device
Manager
The thin layer detection is initially enabled. It can be set to enabled or disabled.
Prerequisites
The thin layer detection is only applicable to:
• Firmware version 2.L3 or later.
• Measurement mode: Level and Interface Level.
• Probe types: Single lead or Coaxial/Large coaxial (standard/HP/C).
Note
When the thin layer detection is disabled, the minimum detectable upper product
thickness is increased, typically 5.1 in. (13 cm).
Procedure
1. HART
®
: Select Configure → Manual Setup → Level Setup.
FOUNDATION
™
Fieldbus: Select Configure → Manual Setup.
2. Select the Environment tab.
The Use Thin Layer Detection check box displays the current state of the thin layer
detection, enabled or disabled.
3. Click the Thin Layer Detection button.
4. Select desired option to enable or disable the thin layer detection, and then select
Next.
5. Select Finish when the Method Complete message appears.
D.9.3
Enable/disable thin layer detection using handheld
communicator
The thin layer detection is initially turned on. It can be set to on or off.
Prerequisites
The thin layer detection is only applicable to:
• Firmware version 2.L3 or later.
• Measurement mode: Level and Interface Level.
Reference Manual Advanced configuration
00809-0100-4530 April 2019
Reference Manual 419