94
Reference Manual
00809-0100-4026, Rev HA
Section 5: Basic Configuration/Start-up
November 2014
Basic Configuration/Start-up
Tank Geometry
6. Select the Tank Type corresponding to the actual
tank. If none of the available options matches
the tank, select Unknown.
HART command: [2, 1, 2, 1].
F
OUNDATION fieldbus parameter:
TRANSDUCER_1100 > GEOM_TANK_TYPE.
Tank Bottom Type is important for the
measurement performance close to the tank
bottom.
HART command: [2, 1, 2, 2].
F
OUNDATION fieldbus parameter:
TRANSDUCER_1100 >
GEOM_TANK_BOTTOM_TYPE.
Tank Height is the distance from the Upper
Reference Point to the Lower Reference Point
(see “Tank geometry” on page 75). This number
needs to be as accurate as possible.
HART command: [2, 1, 2, 3].
F
OUNDATION fieldbus parameter:
TRANSDUCER_1100 > GEOM_TANK_HEIGHT.
Select the Enable Still-Pipe/Bridle
Measurement check-box and enter the Pipe
Inner Diameter if the transmitter is mounted on a
Still-Pipe.
HART command: [2, 1, 2, 4] / [2, 1, 2, 5].
F
OUNDATION fieldbus parameter:
TRANSDUCER_1100>
SIGNAL_PROC_CONFIG (Enable),
ANTENNA_PIPE_DIAM.
For more information, see “Tank geometry” on
page 75.
Enable Still
Pipe/Bridle
Measurement