Cerabar S/Deltabar S Description of parameters (ToF Tool)
Endress+Hauser 111
ZERO POSITION
Entry
Enter value for level offset.
If the measurement should not start at the mounting location of the device, e.g. for
containers with a sump, carry out zero point shift (level offset).
Prerequisite:
• CALIBRATION MODE = dry
Factory setting:
0.0
P01-PMP75xxx-19-xx-xx-xx-001
Fig. 25: Zero point shift
1 Device is mounted above the level lower range value: a positive value has to be
entered for ZERO POSITION.
2 Device is mounted below the level lower range value: a negative value has to be
entered for ZERO POSITION.
DAMPING VALUE
Entry
Enter damping time (time constant τ).
The damping affects the speed at which all subsequent elements, such as the on-site
display, measured value and OUT value of the Analog Input Block react to a change in the
pressure.
Input range:
0.0...999.0 s
Factory setting:
2.0 s or as per order specifications
ToF Tool, Table 7: OPERATING MENU → SETTINGS → BASIC SETUP "Level", LEVEL MODE "Linear"
Parameter name Description
➀
➁