EasyManua.ls Logo

Endress+Hauser cerabar S - Flow Measurement (Tof Tool); Calibration

Endress+Hauser cerabar S
172 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cerabar S/Deltabar S Flow measurement (ToF Tool)
Endress+Hauser 89
9 Flow measurement (ToF Tool)
9.1 Calibration
Example:
In this example, a volume flow should be measured in m
3
/h.
!
Note!
The "Flow measurement" measuring mode is only available for the Deltabar S differential pressure
transmitter.
See also Operating Instructions BA301P Deltabar S, Section 6.5 "Flow measurement".
For a description of the parameters mentioned, see
Page 98, ToF Tool, table 4: POSITION ADJUSTMENT
Page 125 ff, ToF Tool, table 10: BASIC SETUP
For a description of further parameters, see
Page 130, ToF Tool, table 13: EXTENDED SETUP
Page 144, ToF Tool, table 23: PROCESS VALUES.
Description
1 Before configuring the device for your application, the
pressure piping must be cleaned and the device filled
with fluid. See Operating Instructions BA301P, Section
6.5.1.
P01-xMD7xxx-05-xx-xx-xx-010
Fig. 17: Flow measurement calibration
1 See table, step 7.
2 See table, step 8.
2 Carry out position adjustment if necessary. See Page 98,
ToF Tool, table 4: POSITION ADJUSTMENT.
3 If necessary, select the "Flow" measuring mode via the
MEASURING MODE parameter.
Menu path: OPERATING MENU SETTINGS
BASIC SETUP MEASURING MODE
4 Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
5 Select the "Volume p. cond." option by means of the
FLOW-MEAS. TYPE parameter.
6 Select a flow unit via the UNIT FLOW parameter, here
m
3
/h for example.
7Select MAX. FLOW parameter.
Enter the maximum flow value of the primary device,
here 6000 m
3
/h for example. See also layout sheet of
primary device.
8 Select MAX PRESS. FLOW parameter.
Enter the maximum pressure, here 400 mbar for
example. See also layout sheet of primary device.
9 If necessary, use the Analog Input Block to configure the
CHANNEL parameter ( Page 67), L_TYPE parameter
( Page 69), XD_SCALE parameter ( Page 73) and
OUT_SCALE parameter ( Page 71).
10 Result:
The device is configured for flow measurement.
p
[mbar]
6000
0
0 400
[m /h]
3
V

Related product manuals