EasyManua.ls Logo

Endress+Hauser cerabar S - Totalizers

Endress+Hauser cerabar S
172 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cerabar S/Deltabar S Flow measurement (ToF Tool)
Endress+Hauser 91
9.2 Totalizers
Example:
In this example, the volume flow should be totalised and displayed in the unit m
3
E
3
. Negative flows
should be added to the flow rate.
!
Note!
For a description of the parameters mentioned, see
Page 134 ff, ToF Tool, table 15: TOTALIZER SETUP
Page 144 ff, ToF Tool, table 23: PROCESS VALUES
Totalizer 1 can be reset. Totalizer 2 cannot be reset.
!
Note!
You can also specify a customer-specific unit. See parameter description for TOTALIZER 1
UNIT (Page 134) and TOTALIZER 2 UNIT (Page 135).
The TOTALIZER 1 and TOTAL. 1 OVERFLOW parameters display the totalised flow value of the
first totalizer. The TOTALIZER 2 and TOTAL. 2 OVERFLOW parameters display the totalised
flow value of the second totalizer. See Page 144 ff, PROCESS VALUES function group.
You can use the MENU DESCRIPTOR parameter (Page 136) to specify which measured value
should be displayed on the on-site display.
Description
1 Calibrate the device as per Section 5.1.
2 Change the function group:
OPERATING MENU SETTINGS TOTALIZER
SETUP
3 Select a flow unit via the TOTALIZER 1 UNIT
parameter, here m
3
E
3
for example.
4 Use the NEG. FLOW TOT. 1 parameter to specify the
totalising mode for negative flows, here the "Positive"
option for example.
5 Reset totalizer 1 to zero via the RESET TOTALIZER
parameter.
6Result:
The TOTALIZER 1 and TOTAL. 1 OVERFLOW
parameters display the totalised volume flow.

Related product manuals