Installation Gammapilot M FMG60 with FOUNDATION Fieldbus
20 Endress+Hauser
3.3 Water cooling
For the versions of the Gammapilot with water cooling jacket, the following applies:
Material: 316L
Water connection: 2 x G 1/4"A, DIN ISO 228
Inlet temperature: max. 40 °C
Outlet temperature: max. 50 °C (temperature monitoring recommended)
Water pressure: 4 to 6 bar
L00-FMG60xxx-11-00-00-xx-001
Connections of the water cooling jacket
"
Caution!
It is recommended to ground the water cooling jacket separately at the provided earth terminal
(see above picture)
The ambient temperature of the compartment housing must not exceed 75 °C. This is also valid,
if water cooling is applied.
3.3.1 Mounting versions
L00-FMG60xxx-17-00-00-xx-009
A: Recommended mounting position for level measurement: compartment housing at the bottom;
B: In exceptional cases (e.g. shortage of space) the compartment housing may be located at the top.
C: Mounting position for level limit detection and density measurement.
"
Caution!
The water inlet must always be at the bottom to ensure that the water cooling jacket is completely
filled.
T < 75 °C
T < 120 °C
G1/4"AG1/4"A
*
The three screws, connecting the pipe
housing to the compartment housing, must
not be opened.
L00-FMG60xxx-05-00-00-xx-052