Gammapilot M FMG60 with FOUNDATION Fieldbus Accessories
Endress+Hauser 91
8.4 Mounting device FHG60 (for level measurement and
level limit measurement)
8.4.1 Dimensions
L00-FMG60xxx-17-00-00-xx-006A
1a: with preferred mounting method "housing head up" - bracket (for "level" application only
10))
); 1b: with alternative
mounting method "housing head down" - bracket (for "level" application only
1)
); 2: Mounting clamps (number depending
on selected application); 3: Retainers (number and size depending on selected application
10)
; Allen screws to ISO4762
are supplied); 4: Spacing as large as possible
Size of the retainers (depending on the selected application):
"
Caution!
Max. torque for the screws of the retainers: 6 Nm.
88
35
A
B
40
70
70
190
63
35
110
230
Ø 40-65
88
116
M6x25-A2
M6x70-A2
1a
1b
ØC
#
max. 6 Nm
#
max. 6 Nm
2
3
4
10) See "Use" and "ordering information for complete mounting set".
Mounting position at the
FMG60
Length A [mm] Width B [mm] C [mm] Arrangement
(see Fig. overleaf)
Scintillator pipe dimensions 196 126 80 (a)
Electronic pipe dimensions 210 150 102 (b)
Water cooling jacket dimensions 230 200 140 (c)