Commissioning Gammapilot M FMG60 with FOUNDATION Fieldbus
86 Endress+Hauser
6.5.4 Linearization (for concentration measurements)
If the concentration is to be measured in a unit other than the "density unit" (*06), a linearization
must be performed after the calibration. This can be done in the "Linearization" (*4) function
group. The functions of this group and the linearization procedure are described in Operating
Instructions BA 287F, "Gammapilot M - Description of Device Functions". This document can be
found on the CD-ROM supplied for the "ToF Tool - FieldTool Package".
6.5.5 Additional settings
After the calibration, the Gammapilot M outputs the measured value via the PROFIBUS PA
interface.
Many additional functions are available for optimizing the measuring point. They can be configured
as required. A detailed description of all instrument functions is given in Operating Instructions
BA287F, "Gammapilot M - Description of Device Functions". This document can be found on the
CD-ROM supplied for the "ToF Tool - FieldTool Package/FieldCare".
6.6 Density measurement/temperature-compensated
Perform density measurement as described in the section "Calibration for density and concentration
measurements" and then perform temperature calibration (see "Temperature compensation" section
in function description "BA287F/00".
6.7 Gammagraphy detection
See "Gammagraphy" section in function description "BA287F/00".