Gammapilot M FMG60 with FOUNDATION Fieldbus Commissioning
Endress+Hauser 63
6.3 Basic setup
6.3.1 "Present date" (*01)
Date and time of the basic setup are specified in this function.
Format:
DD.MM.YY HH:MM
where:
DD: the day ("01" to "31")
MM: the month ("01"=January to "12"=December)
yy: the year (e.g. "04" for 2004)
HH: the hour ("00" to "23")
MM: the minute ("00" to "59")
Each of these values must be confirmed by "E" after it has been entered.
6.3.2 "beam type" (*02)
This function is used to specify whether the radiation source used emits radiation continuously or
whether it is modulated (for gammagraphy suppression).
Options:
standard/continuous (permanent, continuous radiation)
modulated (modulated radiation source) - under development
6.3.3 "Isotope" (*03)
This function is used to specify which isotope is used for the measurement.
The Gammapilot M needs this information for the decay compensation.
Options:
137
Cs
60
Co
no compensation
present date *01
17.11.04 10:30
dd.mm.yy hh:mm
ENDRESS + HAUSER
E
+
beam type *02
ÂStandard/cont.
Modulated
ENDRESS + HAUSER
E
+
isotope *03
Â137Cs
60Co
no compens.
ENDRESS + HAUSER
E
+