Gammapilot M FMG60 with FOUNDATION Fieldbus Wiring
Endress+Hauser 27
4.9 Wiring in terminal compartment B
"
Caution!
Before connection please note the following:
Connect potential matching line to
transmitter ground terminal and to the ground
terminal of the water cooling jacket (if
present) before connecting up the device (see
above, section "Potential equalization").
The cable isolations must comply with the
supply voltage and the overvoltage category.
The temperature resistance of the connecting
cable must comply with the ambient
temperature.
The procedure
1. Unscrew the cover of the terminal
compartment.
2. Push the following cables through the
appropriate cable glands or threads:
signal cable (if the signal output is located
in terminal compartment B)
PT-100 cable (if present)
cascading cables (input and/or output, if
required)
3. Wire up according to the terminal
assignment diagram.
4. Tighten the cable glands or threads.
5. Screw the cover securely back onto the
terminal compartment.
L00-FMG60xxx-04-00-00-xx-006
1
2
3
4
5
+
-
PT100
IN CASCADE OUT
SIM WP
OFF
ON
3
4
5
6
7
8
9
10
11
12
+ - - +
0
13
B