Gammapilot M FMG60 with FOUNDATION Fieldbus Commissioning
Endress+Hauser 85
"Reference pulse rate" (*1F)
This function displays the reference pulse rate I
0
, which results from the currently active calibration
points. The value cannot be edited.
!
Note!
I
0
is the pulse rate for the tube being empty (theoretical reference value). Generally, the value is
significantly higher than any real pulse rate occurring during the measurement.
"Next point" (*19)
This function is used to specify, if a further calibration point is to be entered or not.
Options:
no
This option must be selected if no further calibration point is to be entered or changed. After this
selection the Gammapilot M returns to the group selection and the calibration is completed.
yes
This option must be selected if a further calibration point is to be entered or changed. The
Gammapilot M returns to the "calibr. point" (*1A) function and the next point can be entered
or changed.
ref. pulse rate *1F
31687 cps
ENDRESS + HAUSER
E
+
next point *19
Âno
yes
ENDRESS + HAUSER
E
+