Commissioning Gammapilot M FMG60 with FOUNDATION Fieldbus
80 Endress+Hauser
6.5.2 Background calibration
Excerpt from the operating menu
The following excerpt from the operating menu shows how the background calibration is entered.
The individual functions are explained in the sections below.
L00-FMG60xxx-05-00-00-en-044
"Background calibration" (*10)
This function is used to start the background calibration
Options:
stop/edit This option must be selected if
no background calibration is to be performed but the pulse rate of an existing background
calibration is to be displayed instead.
a manual background calibration is to be performed.
After selecting this option, the Gammapilot M changes to the "bgr. pulse rate" (*12) function,
where the existing pulse rate is displayed and can be changed if required.
start
This option is used to start an automatic background calibration. The Gammapilot M changes to
the "avg. pulse rate" (*11) function.
*10
backgr. calib.
0:stop/edit
1:start
*11
avg. pulse rate
*12
bgr. pulse rate
-1:not calibrated
1
E
0
E
backgr. cal. *10
stop/edit
start
ENDRESS + HAUSER
E
+