Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54
150 Endress+Hauser
Prerequisite
Only visible for Switch output function = Limit and Assign limit ¹ Off.
Description
Defines the delay for the switching on of the output.
Range of values
0 to 100 s
Factory setting
0 s
Switch-off delay
Navigation
Setup ® Advanced setup ® Switch output ® Switch-off delay
Prerequisite
Only visible for Switch output function = Limit and Assign limit ¹ Off.
Description
Defines the delay for the switching off of the output.
Range of values
0 to 100 s
Factory setting
0 s
Switch output failure mode
Navigation
Setup ® Advanced setup ® Switch output ® Switch output failure mode
Description
Defines the switching state of the output in the case of an error.
Options
• Actual status
The switch output remains in the state it was in when the error occurred
•
Open
• Closed
Factory setting
Open
Switch status
Navigation
Setup ® Advanced setup ® Switch output ® Switch status
Diagnostics ® Measured value ® Switch status
Description
Indicates the current state of the switch output.
Display options
• Open
• Closed