Micropilot FMR60B PROFINET with Ethernet-APL Technical data
Endress+Hauser 81
Plastic housing; process temperature –40 to +130 °C (–40 to +266 °F)
A0032024
38 Plastic housing; process temperature –40 to +130 °C (–40 to +266 °F)
P1 = T
p
: –40 °C (–40 °F) | T
a
: +76 °C (+169 °F)
P2 = T
p
: +76 °C (+169 °F) | T
a
: +76 °C (+169 °F)
P3 = T
p
: +130 °C (+266 °F) | T
a
: +41 °C (+106 °F)
P4 = T
p
: +130 °C (+266 °F) | T
a
: –40 °C (–40 °F)
P5 = T
p
: –40 °C (–40 °F) | T
a
: –40 °C (–40 °F)
In the case of devices with a plastic housing and CSA C/US approval, the selected
process temperature of –40 to +130 °C (–40 to +266 °F) is limited to
0 to +130 °C (+32 to +266 °F).
Restriction to a process temperature of 0 to +130 °C (+32 to +266 °F) with CSA C/US
approval and plastic housing
A0048826
39 Plastic housing; process temperature 0 to +130 °C (+32 to +266 °F) with CSA C/US approval
P1 = T
p
: 0 °C (+32 °F) | T
a
: +76 °C (+169 °F)
P2 = T
p
: +76 °C (+169 °F) | T
a
: +76 °C (+169 °F)
P3 = T
p
: +130 °C (+266 °F) | T
a
: +41 °C (+106 °F)
P4 = T
p
: +130 °C (+266 °F) | T
a
: 0 °C (+32 °F)
P5 = T
p
: 0 °C (+32 °F) | T
a
: 0 °C (+32 °F)
Plastic housing; process temperature –40 to +150 °C (–40 to +302 °F)
A0032024
40 Plastic housing; process temperature –40 to +150 °C (–40 to +302 °F)
P1 = T
p
: –40 °C (–40 °F) | T
a
: +76 °C (+169 °F)
P2 = T
p
: +76 °C (+169 °F) | T
a
: +76 °C (+169 °F)
P3 = T
p
: +150 °C (+302 °F) | T
a
: +25 °C (+77 °F)
P4 = T
p
: +150 °C (+302 °F) | T
a
: –40 °C (–40 °F)
P5 = T
p
: –40 °C (–40 °F) | T
a
: –40 °C (–40 °F)
In the case of devices with a plastic housing and CSA C/US approval, the selected
process temperature of –40 to +150 °C (–40 to +302 °F) is limited to
0 to +150 °C (+32 to +302 °F).