Description of device functions Proline Prowirl 73
162 Endress+Hauser
SAVE CHANGES Description
Use this function to select how the entries in the gas group are saved and used for flow
measurement.
Options
CANCEL
The modified parameters are saved in the gas group but are not used to calculate the
flow.
The gas group can be reactivated at a later stage. To do so, call up the gas group, check
the entries and activate the group by selecting YES.
YES
The modified parameters are saved in the gas group and are used to calculate the flow.
DISCARD
The modified parameters are not saved. The previous values remain valid and are used to
calculate the flow.
Function description, GAS MIXTURE group