Proline Prowirl 73 Technical data
Endress+Hauser 83
< 1.7% o.f.s (2.0% for remote version) for Re between 4000 and 20 000 and
process pressure < 40 bar abs (580 psi abs)
< 2.6% o.r. (2.9% o.r. for remote version) for Re > 20 000 and
process pressure < 120 bar abs (1740 psi abs)
< 2.6% o.f.s (2.9% o.r. for remote version) for Re between 4000 and 20 000 and
process pressure < 120 bar abs (1740 psi abs)
Mass flow (water):
< 0.85% o.r. (1.15% o.r. for remote version) for Re > 20 000
< 0.85% o.f.s (1.15% o.f.s for remote version) for Re between 4000 and 20 000
Mass flow (customer-defined liquids):
To specify the system accuracy, Endress+Hauser requires information on the type of liquid and its
operating temperature, or information in tabular form on the dependency between the liquid
density and temperature.
Example:
Acetone is to be measured at fluid temperatures between 70 and 90 °C. The parameters
TEMPERATURE VALUE (here 80 °C), DENSITY VALUE (here 720.00 kg/m3) and EXPANSION
COEFFICIENT (here 18.0298 × 10E-4 1/°C) have to be entered in the transmitter for this
purpose. The overall system uncertainty, which is smaller than 0.9% for the example cited above,
is made up of the following measuring uncertainties: Uncertainty of volume flow measurement,
uncertainty of temperature measurement, uncertainty of the density-temperature correlation
used (incl. the resulting uncertainty of density).
Mass flow (other fluids):
Depends on the pressure value specified in the OPERATING PRESSURE function (ä 150). An
individual error observation must be carried out.
o.r. = of reading, o.f.s = of full scale value, Re = Reynolds number
Diameter mismatch correction
In the Prowirl 73 shifts in the calibration factor which are caused by a diameter mismatch between
the device and the mating pipe can be corrected (ä 140). The diameter mismatch should only
be corrected within the limit values listed below for which test measurements have also been
performed.
Flange connection:
DN 15 (½"): ±20% of the internal diameter
DN 25 (1"): ±15% of the internal diameter
DN 40 (1½"): ±12% of the internal diameter
•DN 50 ( 2"): ±10% of the internal diameter
Wafer:
DN 15 (½"): ±15% of the internal diameter
DN 25 (1"): ±12% of the internal diameter
DN 40 (1½"): ±9% of the internal diameter
•DN 50 ( 2"): ±8% of the internal diameter
Repeatability ±0.25% o.r. (of reading)
Reaction time/
step response time
If all configurable functions are set to 0, you must reckon with a reaction time/step response time
of 200 ms for vortex frequencies as of 10 Hz. Other settings require a reaction time/step response
time of 100 ms to be added to the total filter reaction time for vortex frequencies as of 10 Hz.
FLOW DAMPING ä 170
DISPLAY DAMPING ä 114
TIME CONSTANT (current output) ä 120
TIME CONSTANT (status output) ä 134