EasyManua.ls Logo

Endress+Hauser Waterpilot FMX21 - Measuring Mode Selection; Selecting the Language

Endress+Hauser Waterpilot FMX21
135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Waterpilot FMX21 Commissioning
Endress+Hauser 39
9.4 Selecting the language
9.4.1 Configure language
A0017654-EN
9.5 Measuring mode selection
The device is configured for the "Pressure" measuring mode as standard. The
measuring range and the unit in which the measured value is transmitted correspond
to the data on the nameplate.
L
WARNING
Changing the measuring mode affects the span (URV)
This situation can result in product overflow.
If the measuring mode is changed, the setting for the span (URV) must be checked in
the "Setup" operating menu and readjusted if necessary.
Measuring mode
Navigation Setup → Measuring mode
Write permission Operators/Service engineers/Expert
Description Select the measuring mode.
The operating menu is structured differently depending on the measuring mode selected.
Options Meas.
Level
Factory setting Level