EasyManuals Logo

Excell PW User Manual

Excell PW
58 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #41 background imageLoading...
Page #41 background image
EXCELL PRECISION CO., LTD
PW Service Manual SME300000016
39
3-3-5 Noise Filter Settings
Select in the general function setting mode to set the
noise filter settings.
4 When modifying , the parameters of remain un-altered.
4 In approved models, set as 1 or 3, setting is not available
Returning to the zero point setting
Use
ON/ZERO
or
UNITS
key to key in the
parameters or zero point
Default setting = 0
0 No skip 5 skip 5d
1 skip 1d 6 skip 6d
2 skip 2d 7 skip 7d
3 skip 3d 8 skip 8d
4 skip 4d 9 skip 9d
n When the weight on the scale is over 1/3
full capacity, the function is on.
Digital switch & Stabilization range
setting
LCD displays the last parameter setting
Filter parameter setting
LCD displays the last parameter setting
Returning to zero point setting
LCD displays the last status
Filter parameter setting
Use
ON/ZERO
or
UNITS
keys to key in the
parameters. Default setting = 5
Parameter 0 ~ 9.
The larger the number is, the faster the
filter responds
. Fast response can lead to
weigh
ing instability.
Digital switch & Stabilization range
setting
Use
ON/ZERO
or
UNITS
keys to key in the
parameters. Default setting = 0
Parameter 0 ~ 9, the larger the number
the
more stable the weight.
ACTIONS
DISPLAY
NOTE
Press
TARE
key
Use
ON/ZERO
or
UNITS
key to key in the
parameters
Press
TARE
key
Use
ON/ZERO
or
UNITS
key to key in the
parameters
Press
TARE
key
Use
ON/ZERO
or
UNITS
key to key in the
parameters
Press
TARE
key

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Excell PW and is the answer not in the manual?

Excell PW Specifications

General IconGeneral
BrandExcell
ModelPW
CategoryAccessories
LanguageEnglish

Related product manuals