3-3-5 Noise Filter Settings
Select in the general function setting mode to set the
noise filter settings.
4 When modifying , the parameters of remain un-altered.
4 In approved models, set as 1 or 3, setting is not available
Returning to the zero point setting
Use
ON/ZERO
or
UNITS
key to key in the
parameters or zero point
Default setting = 0
0 ⇒ No skip 5 ⇒ skip 5d
1 ⇒ skip 1d 6 ⇒ skip 6d
2 ⇒ skip 2d 7 ⇒ skip 7d
3 ⇒ skip 3d 8 ⇒ skip 8d
4 ⇒ skip 4d 9 ⇒ skip 9d
n When the weight on the scale is over 1/3
full capacity, the function is on.
Digital switch & Stabilization range
setting
LCD displays the last parameter setting
Filter parameter setting
LCD displays the last parameter setting
Returning to zero point setting
LCD displays the last status
Filter parameter setting
Use
ON/ZERO
or
UNITS
keys to key in the
parameters. Default setting = 5
Parameter 0 ~ 9.
The larger the number is, the faster the
filter responds
. Fast response can lead to
ing instability.
Digital switch & Stabilization range
setting
Use
ON/ZERO
or
UNITS
keys to key in the
parameters. Default setting = 0
Parameter 0 ~ 9, the larger the number
ACTIONS
DISPLAY
NOTE