3-3-6 Hold Function Settings
Select in the general function setting mode to set the
hold function.
ON/ZERO
key ⇒ Increment number by 1
UNITS
key ⇒ Decrement number by 1
TARE
key ⇒ Move cursor one place to the right
NET/GROSS
key ⇒ Move cursor one place to the left
Hold function setting
Use
ON/ZERO
or
UNITS
keys to key in the
parameters
Default setting = 0
0 ⇒ Hold function disabled
1 ⇒ “Peak hold” mode
2 ⇒ “Stable hold 1” mode
3 ⇒ “Stable hold 2” mode
Hold function setting
LCD displays the last parameter setting
Press
key
Use
ON/ZERO
or
UNITS
key to key in the
parameters
Press
TARE
key
Key in the parameter “00”
Press
TARE
key to
complete the general
setting procedures
ACTIONS
DISPLAY
NOTE
= Hold function disabled
= “Peak hold” mode
Keep displaying the maximum weight when the weight is continually changing
To exit this mode, press any key.
= “Stable hold 1” mode
When the weight is stable, the LCD shows the current weight value. To exit
this mode, press any key.
= “Stable hold 2” mode
When the weight is stable, the LCD shows the current weight value. When the
weight returns to zero (<10d), the hold mode is cancelled automatically.