EXCELL PRECISION CO., LTD
PW Service Manual SME300000016
51
3-5-8 Output Condition Settings
Select in the RS232 serial interface setting mode to
Reset of Auto Transmission.
ON/ZERO
key Increment number by 1
UNITS
key Decrement number by 1
TARE
key Move cursor one place to the right
NET/GROSS
key Move cursor one place to the left
Output condition settings
The LCD shows the last setting.
n Default setting:
Output condition settings
Use
ON/ZERO
or
UNITS
keys to select the
desired setting:
= Output all the time
= Stable weight output only (No
output when under/overload or unstable)
Press
TARE
key
Use
ON/ZERO
or
UNITS
keys to select the desired
setting
Press
TARE
key
Key in the parameter 00
Press the
TARE
key to exit
RS232 serial interface
settings
ACTIONS
DISPLAY
NOTE