What to do if there is low photon flux in Excillum Laboratory Equipment?
- GGregory WardNov 18, 2025
If you're experiencing low photon flux relative to electron-beam power in your Excillum Laboratory Equipment, there are several potential causes. One reason could be that the electron beam isn't hitting the metal jet in the optimal position; try adjusting the electron-beam position relative to the metal jet. Another cause might be a contaminated or out-of-order exit window; check the exit window heating current and voltage – if it's not approximately 1.5 A and 4 V, replace the exit window according to the service manual instructions. A large leakage current in electron caused by contamination of high voltage feedthrough, metal electrodes and vacuum chamber, a faulty high-voltage cable or a faulty high-voltage power supply unit may be the reason. In this case, immediately contact Excillum a...