IEEE 488.2 and Specific Command Reference
FTB-2/FTB-2 Pro and FTB-4 Pro 405
IEEE 488.2 Required Commands
*ESE
Description The *ESE command sets the Standard Event
Status Enable Register bits, as defined in the
table below. This register contains a mask value
for the bits to be enabled in the Standard Event
Status Register.
Syntax *ESE<wsp><RegisterValue>
Parameter(s) RegisterValue:
The program data syntax for <RegisterValue> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
The <RegisterValue>, expressed in base 2,
represents the bit values of the Standard Event
Status Enable Register.
PON URQ CME EXE DDE QYE N.U. OPC
MSB LSBStandard Event Status Enable Register