IEEE 488.2 and Specific Command Reference
FTB-2/FTB-2 Pro and FTB-4 Pro 435
Specific Commands
:SYSTem:ERRor[:NEXT]?
Description The SYSTem:ERRor[:NEXT]? queries the
error/event queue for the next item and removes
it from the queue. The response message
consists of two fields separated by commas
<Code>,<Description[,Info]>.
SYSTem:ERRor[:NEXT]? is a query only and,
therefore, does not have an associated *RST
state.
Syntax :SYSTem:ERRor[:NEXT]?
Parameter(s) None
Response Syntax <Code>,<Description[,Info]>
Response(s)
Code:
The response data syntax for <Code> is defined
as a <NR1 NUMERIC RESPONSE DATA>
element.
The <Code> is a unique integer in the range
[-32768, 32767]. All positive numbers are
instrument-dependent. All negative numbers are
reserved by the SCPI standard with certain
standard error/event codes described in an
appendix of this document. The zero value is
also used to indicate that no error or event has
occurred.
Description[,Info]:
The response data syntax for
<Description[,Info]> is defined as a <STRING
RESPONSE DATA> element.
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com

Table of Contents

Other manuals for EXFO FTB-2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EXFO FTB-2 and is the answer not in the manual?

EXFO FTB-2 Specifications

General IconGeneral
BrandEXFO
ModelFTB-2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals