FM5000C - FM5000PLUS
Maternal/Fetal Monitor
Harmful ingress protection of
transducers
(7) ECG
3 leads, 5 leads, 12 leads; auto recognition
12 leads (R, L, F, N, C; or RA, LA, LL, RL, V)-(R, L, F, N, C1, C2, C3, C4, C5,
C6 or RA, LA, LL, RL, V1, V2, V3, V4, V5, V6)
I, II, III, avR, avL, avF, V- V1, V2, V3, V4, V5, V6
5 leads (R, L, F, N, C or RA, LA, LL, RL, V)
I, II, III--I, II, III, avR, avL, avF, V
3 leads (R, L, F or RA, LA, LL)
Sensitivity (Gain) and
error
2.5mm/mV(×0.25), 5mm/mV(×0.5), 10 mm/mV(×1), 20mm/mV(×2) and
auto; when applying ±300mV DC polarity voltage, the error is ±5%.
Frequency
characteristics
(Display and
recording)
Surgical mode: 1 Hz~20 Hz (-3.0dB~+0.4dB);
Monitor mode: 0.5 Hz~40 Hz (-3.0dB~+0.4dB);
Diagnostic mode: 0.05Hz~130 Hz (-3.0dB~+0.4dB);
Monitor and surgical mode: the 50/60Hz notch function is auto activated;
Diagnostic mode: the 50/60Hz notch function can be select to activate or
deactivate mannually.
Power frequency interference rejection capability≥20dB
Load 1V, power frequency, differentialmode AC voltage for 10s without
damage (p-v)
Respiration, lead-off,
sensing and active
noise suppression
QRS wave amplitude
and interval
Width (pediatric/neonate)
No response to the signals:
a) with amplitude not exceeding 0.15mV (p-v
RTI) (except for pediatric/neonate mode); or
b)with 10ms width (except for pediatric/neonate
mode) in case of 1mV amplitude.
Triangular wave amplitude (p-v RTI)
QRS wave amplitude (p-v RTI)
QRS wave recurrence frequency
HR measurement
range and accuracy
± 1% or ±1bpm, whichever is the
larger