II-4
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
HR Alarm range
Adult
15~300bpm
Pediatric/Neonate
15~350bpm
Alarm error
±1bpm
Alarm start time for
asystole
<10s
Alarm start time for
heart rate too low or
too high
<10s
Input dynamic range
Input signal amplitude
±5mV
Rate (RTI)
320mV/s
DC offset voltage
-300 ~ +300mV
Output signal change
±10%
Failure display (attenuation
prior to display)
Not reduced to below 50%
Input impedance
5M
System noise (p-v
RTI)
<30µV
Multi-channel
crosstalk
<5%
Time reference
selection and accuracy
Time reference
selection
Permanent display
25mm/s
Non-permanent display
12.5mm/s, 25mm/s, 50mm/s
Accuracy
±10%
Output display
Channel width
30mm
Aspect ratio
0.4s/mV
Input signal
reproduction accuracy
Total system error
±20% or ±100µV, whichever is larger
Frequency response
Sinusoidal input
0.67~40Hz
(attenuation-3dB)
Response to 20ms
(width) triangular wave
0~25Hz attenuation in
amplitude of wave peak
Response to 0.3 mV·s
shock in the shock
range
Offset (RTI)
0.1mV
Slope (RTI)
≤0.30mV/s
Electrode weighting
factor
≥±5%
Hysteresis effect of
15mm offset
≤0.5mm
Calibration voltage
1mV , error: ±5%
Common mode
rejection
<1mV (p-v RTI)