5. Diagnosis and error treatment
5-5
Festo P.BE-SPC11-SYS-ASI-EN en 0203NH
Fault no. Eliminating faultsOperating status
Amplification stage, cushioning stage or
system parameter set incorrectly.
S Correct the parameters
The measuring system has no valid length
code (non-permitted measuring system
length or length code does not exist).
Ser vicing required
SPC11 damaged ( E EPROM-ACK fault). S Replace the SPC11
Measuring system type is not supported
(invalid measuring system ID read).
S Replace the measuring system
Communication to T emposonics is faulty
(non-plausible data has been received).
S Check the measuring system
length or replace the measuring
system.
Communication to T emposonics is faulty
(non-plausible data has been received).
S Check the measuring system
length or replace the measuring
system.
Incorrect direction of movement ascertained
during positioning stroke; tubing is incor-
rectly connected.
S Correct the tubing of the propor-
tional directional control valve.
Time out, The position c annot be reached
fast enough (positioning timeout approx.
10 s).
S Removeobstructioninpositioning
path or check compressed air
supply.
Intermediate position does not exist or is
outside the permit ted positioning range.
S Increase the positioning range
and teach the end positions
again, in order that the intermedi-
ate positions lie within the per-
mitted positioning range.
or
S Teach the intermediate positions
again.