8 Operator‘s Manual for instrument types MP0R
Measurement
- Measuring mode active
- Measuring mode active
Signal LED /
Display
Acoustic Signals Meaning
green /
reading
1 short signal • Measurement accept.
• Limit monitoring enabled: Reading is
above the set specification limit value.
red /
2 short signals Reading has violated the specification
limit. The captured reading is < 90 % of
the set specification limit value.
Signal LED /
Display
Acoustic Signals Meaning
green /
reading
1 short signal • Measurement accept.
• Limit monitoring enabled: Reading is
between the specification limit values.
red /
2 short signals Reading has violated the upper set
specification limit. The captured read-
ing is > 120 % of the set upper specifi-
cation limit value.
red /
2 short signals Reading has violated the lower set
specification limit. The captured read-
ing is < 80 % of the lower set specifica-
tion limit value.