40 Operator‘s Manual for instrument types MP0R
Instrument Settings
Factory Reset - Menu 9
All settings of all instrument setting menus are set back to the factory
setting
s. A calibration will be deleted, too.
Call up instrument setting menu see description on page 31.
Procedure
1. Calling the instrument setting menu 9, " " will appear on the dis-
play.
2. Press button [OK]. " " will appear on the display.
3. Use the key [] to confirm the reset procedure. " " and " "
appears in succession on the display. The reset of the instrument to
the default factory settings is finished.
Please Note!
The reset to the factory settings deletes also the last calibrations! It is
necessary to perform a calibration on the corresponding reference part for
precise measurement.
Perform a new normalization or corrective calibration before next mea-
surement. Procedure 11.