Operator‘s Manual for instrument types MP0R 7
Measurement
30, Selecting the measuring mode
20, Setting specification limits and an offset value
3.1 Measurement accept
The measurement accept are signaled optical and acoustic. All signals are
listed in the following tables. The set measuring mode will display tempo-
rary after switching on the instrument.
- Measuring mode active
Please Note!
There are no verified instruments available for measuring the coating thick-
ness, as is the case, for example, for weight measurements with balances.
Rather, the measurement results for the coating thickness are influenced by
the materials and the shape of the specimen. However, these individual
influences can taken into account using appropriate calibrations. 10,
"Influencing Variables Taken into Consideration".
Signal LED /
Display
Acoustic Signals Meaning
green /
reading
1 short signal • Measurement accept.
• Limit monitoring enabled: Reading is
between the specification limit values.
red /
2 short signals Reading has violated the upper specifi-
cation limit.
red /
2 short signals Reading has violated the lower specifi-
cation limit.