EasyManua.ls Logo

FISCHER PERMASCOPE MP0R - Lock Function - Menu 1; Block Length - Menu 1

FISCHER PERMASCOPE MP0R
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator‘s Manual for instrument types MP0R 33
Instrument Settings
Lock function - Menu 1
Enable/disable the calibration functions ( and ). This avoids erro-
neou
s measurements due to unintentional adjustments of instrument
parameters.
Call up instrument setting menu 1 see description on page 31.
Block length - Menu 1
Number of single readings collected to a block.
Call up instrument
setting menu see description on page 31.
The mean value of the block is displayed automatically after the set
number of single readings.
Mode Default setting Further settings
Std,
SSPC,
90.10
=
free and free
= locked and locked
= locked and free
Mode Default setting Further settings
Std
, = block creation
disabled
= block creation of n
single readings (n = 2 .... 20)
SSPC
= 3 single readings
collected to a block
= block creation of n
single readings (n = 3 .... 20)
90.10 not available - only manual block creation possible by using the
key [OK]
CAL
O
CAL
O
CAL
O
CAL
O

Table of Contents

Related product manuals