12 Operator‘s Manual for instrument types MP0R
Influencing Variables Taken into Consideration
4.3 Normalization
When necessary
When the readings during a test measurement on a reference part are no
longer acceptable. Test measurements should be carried out for a
measuring application when it has not been used for a while.
Required Materials
Base = part made from substrate material. In geometry and substrate
material, this specimen should correspond exactly to the parts to be
measured. If
• Fe displayed: Utilizes the magnetic induction measuring method.
Base = ferromagnetic reference part without the coating to be measured
from customer's own production.
• NFe displayed: Utilizes the eddy current measuring method.
Base = non-magnetic reference part without the coating to be measured
from customer's own production.
Before you start
Instrument must be turned ON by pressing the key [OK].
For MP0R instrument models: Remove the orange protective cap from
the probe.
Please Note!
Do not use the supplied metal sheet (NF/FE or ISO/NF) for base!!!
As a rule, the material properties will not correspond to the uncoated
specimen from customer‘s own production! Use it only for instrument check!