Operator‘s Manual for instrument types MP0R 13
Influencing Variables Taken into Consideration
Normalization procedure
1. Open the normalization menu: Set the display menu with [ ] or [ ] to
show .
2. Press button [OK]. The message "Base" (for substrate material) ap-
pears.
3. Perform 5 to 10 measurements on the uncoated reference part (=
Base, specimen without the coating). After each measurement, the
current reading will appear on the display.
4. Press button [OK]. The reading 0.00 appears on the display.
The reference point received from the normalization will be stored in
the instrument as zero point.
The instrument is ready to make measurements.
Please Note!
All readings in the memory will be deleted when performing a normalization.
Please Note!
The menu „Normalization“ can be locked to protect the instrument from unin-
tended interventions. Procedure 31 and 33.
Please Note!
If you want to delete wrong readings, press button [ESC]. Then, the
complete data memory will be deleted, but the normalization routine will not
be interrupted. After that, perform five more measurements and go on with
step 4.
Please Note!
Only for DUALSCOPE
®
instrument types: If you want to measure on both
substrate materials a normalization procedure must be carried out for each
substrate material (Fe and NFe).