Operator‘s Manual for instrument types MP0R 17
Influencing Variables Taken into Consideration
4.5 Deleting the Corrective Calibration and Normaliza-
tion / Restoring the baseline of the characteristic
Sometimes, if the instrument does not measure correctly even after the
calibration, the calibration can be deleted. This can occur if the previous
calibration has not been carried out correctly. In such a case, the default
calibration can be restored.
Procedure
Perform a Corrective Calibration modificate as follows:
Differing from the description of the corrective calibration procedure begin-
ning on page 15 use in this special case the supplied metal boards (NF/FE
or ISO/NF) for base. Instead measuring on calibration foil measure on the
supplied metal board again. Set the nominal value to 00.0.
Now the instrument is set to the default calibration.
Please Note!
For DUALSCOPE
®
instrument types: The base line of each of the two
measurement channels (Fe and NFe) must be restored separately.
Please Note!
Perform a new Normalization or Corrective Calibration according the
instruction, beginning on page 10.